34 WIKA operating instructions, models CTH6300 and CTH6500
2079988.04 07/2014 GB/D
GB
7. Interface protocol RS-232
CTH6500
Recognise MEASURED VALUE 1 + 2 automatically FC (hex)
Enable keyboard 0 (hex)
Read version number of rmware I (ASCII)
Read storage (only for instruments with data logger) n (ASCII)
If data are read out from the instrument using the FC (hex) command, the
keyboard is locked. It can be enabled again using the 0 (hex) command.
The data are sent by the instrument in the following format.
Data type = String
The string length depends on whether it is a 1-channel or a 2-channel instrument.
Example: Turbo Pascal source code for interface initialisation
PROCEDURE V24open (com1, com2: Boolean);
{RS 232 initialisation}
VAR value: Byte;
BEGIN
IF com1 THEN BEGIN {of COM 1 initialisation}
PORT [$3FB]: = 128; {Enable Baud entry}
PORT [$3F8]: = 48; {Baud rate - LSB, 2400 Baud}
PORT [$3F9]: = 0; {Baud rate - MSB}
PORT [$3FB]: = 7; {8 bits, No parity, 2 Stop}
PORT [$3FC]: = 2; {DTR = 0, RTS = 1}
Value: = PORT [$3F8] {Clear receiving register}
END;
END;