Wika T53 User Manual Page 21

  • Download
  • Add to my manuals
  • Print
  • Page
    / 26
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 20
21
2.9 PR_CUST_LIN Block (PA Slot 4) Parameter List
2.9.1 Linear interpolation linearisation, Description
LinType 1 = “Linearisation Table” generates a customer specific linear interpolation linearisation. Linear
interpolation linearisation can be used on mV, ohmic and potentiometer signals.The linear interpolation
linearisation is defined by straight lines drawn between the entered X / Y (input / output) coordinates.
The linearisation table must consist of 10 to 50 coordinate sets. The X values of the coordinates must
be entered in ascending order. The lowest and highest X values function as the lower and the upper lim
-
it respectively. All X values must be entered as µV, Ohm or % for Voltage, Resistance or Potentiometer
measurements in that order. The table output will be converted to actual chosen PRMARY_VALUE_UNIT
(Example: 1000 / 3000 as X / Y values: output will read 3,00 if PRIMARY_VALUE_UNIT is set to “mV”
and 1 mV is connected to input).
2.9.2 Linear Interpolation Linearisation, Parameter List.
Parameter
Rel.
Index
PA
Description Type Store
Size
byte
RO /
R/W
Min. Max. Default
TAB_MIN_NUMBER
29 Minimum number of linearisation points allowed (10)
Un-
signed
8
N 1 RO 10
TAB_MAX_NUMBER
30 Maximum number of linearisation points allowed (50)
Un-
signed
8
N 1 RO 50
TAB_ACTUAL_NUMBER
31 Number of linearisation points in the linearisation table.
Un-
signed
8
SRC 1 R/W 11
TAB_X_Y_VALUE1
32 Linearisation x,y coordinate 1
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE2
33 Linearisation x,y coordinate 2
Float
array
SRC 8 R/W
1000,
100
TAB_X_Y_VALUE3
34 Linearisation x,y coordinate 3
Float
array
SRC 8 R/W
2000,
200
TAB_X_Y_VALUE4
35 Linearisation x,y coordinate 4
Float
array
SRC 8 R/W
3000,
300
TAB_X_Y_VALUE5
36 Linearisation x,y coordinate 5
Float
array
SRC 8 R/W
4000,
400
TAB_X_Y_VALUE6
37 Linearisation x,y coordinate 6
Float
array
SRC 8 R/W
5000,
500
TAB_X_Y_VALUE7
38 Linearisation x,y coordinate 7
Float
array
SRC 8 R/W
6000,
600
TAB_X_Y_VALUE8
39 Linearisation x,y coordinate 8
Float
array
SRC 8 R/W
7000,
700
TAB_X_Y_VALUE9
40 Linearisation x,y coordinate 9
Float
array
SRC 8 R/W
8000,
800
TAB_X_Y_VALUE10
41 Linearisation x,y coordinate 10
Float
array
SRC 8 R/W
9000,
900
TAB_X_Y_VALUE11
42 Linearisation x,y coordinate 11
Float
array
SRC 8 R/W
10000,
1000
TAB_X_Y_VALUE12
43 Linearisation x,y coordinate 12
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE13
44 Linearisation x,y coordinate 13
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE14
45 Linearisation x,y coordinate 14
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE15
46 Linearisation x,y coordinate 15
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE16
47 Linearisation x,y coordinate 16
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE17
48 Linearisation x,y coordinate 17
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE18
49 Linearisation x,y coordinate 18
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE19
50 Linearisation x,y coordinate 19
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE20
51 Linearisation x,y coordinate 20
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE21
52 Linearisation x,y coordinate 21
Float
array
SRC 8 R/W 0, 0
TAB_X_Y_VALUE22
53 Linearisation x,y coordinate 22
Float
array
SRC 8 R/W 0, 0