6. Commissioning, operation
pressure gauge, it must be protected against being blocked by debris
and dirt.
With safety pressure gauges (see dial symbol
k
) ensure that the free
space behind the blow-out back is at least 15 mm.
Requirements for the installation point
If the measuring point is not suciently stable a measuring instrument
support such as a bracket or ange should be used for fastening
(possibly via a exible capillary). If vibrations cannot be avoided by
taking suitable measures during installation, liquid-lled instruments
should be used. Instruments must be protected against coarse dirt and
wide uctuations in ambient temperature.
Installation
Nominal position per EN 837-1 / 9.6.7 Figure 9: 90° ( ⊥ )
Process connection lower mount (LM) or back mount (BM)
After mounting, set the compensating valve
(if available) from CLOSE to OPEN.
In order to avoid any additional heating, the instruments must not be
exposed to direct solar irradiation while in operation!