37WIKA operating instructions, models CTH6300 and CTH6500
2079988.04 07/2014 GB/D
GB
9. Faults / 10. Dismounting, return and disposal
Indication Cause Measures
E1oh
Interruption of the Auto-o
function
Switch on the Auto-o function
again, see chapter 6.7 “AUTO-
OFF function”.
E1eh
Interruption of the Auto-o
function
Switch on the Auto-o function
again, see chapter 6.7 “AUTO-
OFF function”.
E23
EE-prom content destroyed Send in for repair
E25
EE-prom content destroyed Send in for repair
E12
Overow Remove battery and insert it
again
E31
Temperature of the reference
measuring point is outside of the
measuring range
Use the temperature probe
according to the technical
specications. See chapter
3 “Specications”.
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the
hand-held thermometer must be shut down immediately.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
10.2 “Return”.
10. Dismounting, return and disposal
WARNING!
Residual media at the dismounted hand-held thermometer and/or
temperature probe can result in a risk to persons, the environment and
equipment.
Take sucient precautionary measures.