Wika A2G-25 User Manual Page 7

  • Download
  • Add to my manuals
  • Print
  • Page
    / 56
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 6
WIKA operating instructions air2guide model A2G-25
7
GB
40202755.02 07/2012 GB/D/F/E/I
3. Specications
Long-term stability
Manual zero adjustment (standard)
At 1,000 and 2,000 Pa: ±8 Pa
At 5,000 and 7,000 Pa: ±24 Pa
Automatic zero adjustment (option)
±1 Pa
Maximum working pressure
25 kPa
Permissible temperature
Ambient: -20 … +70 °C
Operation: -5 … +50 °C
Permissible ambient humidity
0 ... 95 % rH, non-condensing
Ingress protection
IP 54 per EN 60529 / lEC 529
Standards
CE conformity:
2004/108/EG Electromagnetic compatibility (EMC)
2002/95/EC RoHS (restriction of the use of certain hazardous
substances in electrical and electronic equipment)
For further specications see WIKA data sheet SP 69.04 and the order
documentation.