Wika Pascal ET User Manual Page 36

  • Download
  • Add to my manuals
  • Print
  • Page
    / 88
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 35
6 Commissioning, operation
36 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
The instrument requires the channel role:
Figure 24 Channel role selection
NONE to use the channel for simple measurement
REF to use the channel as reference in a comparison calibration
DUT to use the channel as measurement of a Device Under Test in a comparison
calibration
Press one of the above button (NONE, REF, DUT) to define the channel role.
Pressing REF or DUT the procedure continues as shown below:
I/O to assign a particular I/O module to the software channel.
KEYBOARD to setup by keyboard a particular value read from an external instrument
(i.e. analogue gauge).
Figure 25 Channel mode selection
Page view 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 87 88

Comments to this Manuals

No comments