Introduction
This configuration manual
contains the necessary information for configuration of the fieldbus temperature transmit
-
ter T53 via a host system with application software for PROFIBUS
®
(for Foundation Fieldbus
HOST systems please refer to the “Configuration Manual for Foundation Fieldbus”). The au
-
toswitch function of the modules ensures automatic switch to the connected protocol.
The Fieldbus Software
has been developped by WIKA according to the specifications of the Fieldbus Foundation
and the PROFIBUS Nutzerorganisation.
The files for Profibus
®
PA are:
WIKA5310.gsd - Geräte Stamm Datei (Device Description Data)
T53_10D.bmp - Icon for dianostic mode
T53_10N.bmp - Icon for normal mode
T53_10S.bmp - Icon for special operation mode
These files can be downloaded from our homepage www.wika.de.
Please follow the instructions for the application software in question when installing the
files.
Driver for Siemens PDM software
WIKA has developped Device Description Language files for the Siemens PDM application
software for Profibus
®
PA.
The files for Siemens PDM are:
PA_Sensor_Temp_WIKA_T5310.devices Installation file
WIKA5310.ddl Device Description file
These files can be delivered to interested customers at request.
Please follow the instructions for the Siemens PMD application software when installing the
files.
Parameter lists abbreviations
In the Store column:
SRC = Static Revision Counter; N = No; D = Dynamic;
Cst = Constant. The parameter doesn’t change in a device
In the RO / R/W column:
RO = Read Only; R /W = Read Write; * = Mixed of RO and R/W; ** = Don’t care
1.0 The Physical Block (PA Slot 0), Profibus
1.1 Diagnosis
In order to provide some information about the device to the control application and the hu
-
man interface, there are diagnosis parameters in the device. The diagnosis parameters have
a bit string data type and there is a mask parameter indicating which diagnosis is supported
by the device.
1.2 Diagnosis of the Device Characteristics
In the Physical block the DIAGNOSIS parameter has the information about the “alerts” into
the device (for instance, device not initialized, power up, factory init, hardware failure, etc).
The DIAGNOSIS_MASK has the diagnosis supported by the device.