Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.138
11498171.01 08/2009 GB
After the password has been entered the SETUP Service screen allows access to the
Calibrate, Seal Point, Linerize, and Adaptation screens.
Consult factory before changing any Seal Point, Linerization or Adaptation
parameters.
Press the [CALIBRATE] key to access the calibrate screen, "Figure - Calibrate Data".
The Calibrate screen contains three tabs: DATA, EDIT AND CALIBRATE. When entering the
calibrate screen the first time the Data Screen is the default. The Data screen allows changes
to be made to the Zero, Span, Date of calibration and displays the sensor reading.
The screen accessed by pressing the [EDIT] Tab, "Figure - Calibration Edit", allows calibra-
tion using data available from a previous calibration. An example of this is when an As-Found
calibration is performed and the applied and measured pressures from the calibration are
available. The low true pressure should be less than 20 %FS and the high true pressure
should be greater than 80 %FS for best results. To edit the calibration from known data, enter
the applied pressures in the Desired column and the measured pressures in the Actual column
by pressing the number to be adjusted. When the values are changed, a new key “APPLY” will
appear on the screen. Press the Apply key to save the calibration data.
Warning
Figure - Calibrate Data