Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1 47
11498171.01 08/2009 GB
Command Data Response/Function
Span desired pressure or ? Sets span on active transducer or for
?, clears previous value, must be
> 50 %FS and has a 1 % limit.
Span? <sp>XXXXXXX<cr><lf> Returns span scale factor for active
transducer.
Sparity Even, Odd, None Sets the serial parity.
Sparity? <sp>CCCC<cr><lf> Returns the serial parity.
Sstop 1 or 2 Sets the serial stop bits.
Sstop? <sp>X<cr><lf> Returns the serial stop bits.
Stable? <sp>(Yes or No)<cr><lf> Returns YES if instrument is stable, or
No.
Stabledelay 0 to 65535 Sets the stable time to the number of
seconds specified.
Stabledelay? <sp>XXXXXXX<cr><lf> Returns the stable time.
stabletime 0 to 65535 Sets the stable time to the number of
seconds specified.
Stabletime? <sp>XXXXXXX<cr><lf> Returns the stable time.
StableWin %fs value Sets the stable window as a %FS for
the active sensor and turndown.
StableWin? <sp>XX<cr><lf> Returns the stable window.
Standby None Instrument placed in Standby Mode.
Standby? <sp>(Yes or No)<cr><lf> Returns Yes if instrument is in Standby,
No if otherwise.
Step value inside upper and lower limits
and inside the range of the active
sensor and turndown.
Sets the control step size for the instru-
ment.
Step- Optional value Jogs the setpoint down one step.
Step+ Optional value Jogs the setpoint up one step.
Step? <sp>XXXXXXX<cr><lf> Returns the control step for the instru-
ment.
Step% Value in % of current range Sets the control step in % of current
range.
Steppct Value in % of current range Sets the control step in % of current
range.
Steppct? <sp>n.nnnnne+nn<cr><lf> Returns the current step in % of
current range.
Units units code or text in table below Sets the instrument engineering units.
Units? <sp>CCCC<cr><lf> Returns the instrument units in a text
string.
Unlock None Releases Acquire locks. See Acquire?
and Release?
UpperLimit Value inside primary xducer Range
on turndown #1 in current units.
Sets the Upper control limit for the
active transducer.
UpperLimit? <sp>xxxxxxx<cr><lf> Returns the upper control limit for the
active transducer.