Wika DPT-10 User Manual Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
à Connect minus side always above the max. level
à The condensate vessel ensures a constant pressure on the minus
side
à For measurements in products with solid content such as e.g. dirty
liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
+
min.
max.
1
2
2
3
4
6
5 5
Fig. 18: Measurement setup in closed vessel with superimposed steam
1 Condensate vessel
2 Blocking valves
3 DPT10
4 Precipitator
5 Drain valves
6 3-fold valve block
à Mount DPT10 directly to the vessel
à Connect minus side always above the max. level
à The condensate vessel ensures a constant pressure on the minus
side
à For measurements in products with solid content such as e.g. dirty
liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
In closed vessels with
steam layering with uni-
lateral pressure trans-
mitter
22 Dierential pressure transmitter DPT10 - metal measuring diaphragm F oundation Fieldbus
4 Mounting
37246-EN-100120
Page view 21