Page view 41
6 Adjustment with the indicating and
adjustment module
6.1 Short description
The indicating and adjustment module is used for measured value
display, adjustment and diagnosis. It can be mounted in the following
housing versions and instruments:
l All sensors of the IPT-1* instrument family, in the single as well as
double chamber housing (optionally in the electronics or connec-
tion compartment)
l External indicating and adjustment unit
Note:
You can nd detailed information on adjustment in the operating
instructions manual "Indicating and adjustment module".
6.2 Insert indicating and adjustment module
The indicating and adjustment module can be inserted and removed at
any time. It is not necessary to interrupt the voltage supply.
For installation, proceed as follows:
1 Unscrew the housing cover
2 Place the indicating and adjustment module in the desired position
on the electronics (you can choose any one of four dierent
positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the electronics
and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Function/Conguration
Mount/Dismount indicat-
ing and adjustment
module
42 Dierential pressure transmitter DPT10 - metal measuring diaphragm F oundation Fieldbus
6 Adjustment with the indicating and adjustment module
37246-EN-100120