Operating Instructions Temperature Indicating Controller CS4R
V1.2
•
05/2006 - 13 -
Sensor correction setting
• Sets the sensor correction value of the sensor.
• Thermocouple and RTD inputs : –100.0 … 100.0°C (°F)
DC voltage and current inputs : –1000 … 1000
(The placement of the decimal point follows the selection.)
0.0°C
Communication protocol selection
• Selects communication protocol.
• Available only when the option Digital interface [5] is added.
• WIKA protocol :
Modbus ASCII mode :
Modbus RTU mode :
WIKA protocol
Instrument number setting
• Sets an individual instrument number to each CS4R when
connecting plural CS4R in serial communication.
• Available only when the option Digital interface [5] is added.
• Setting range: 0 … 95
0
Data transfer rate selection
• Selects a data transfer rate in accordance with host computer
• Available only when the option Digital interface [5] is added.
• 2400bps :
4800bps :
9600bps :
19200bps :
9600bps
Parity selection
• Selects the parity.
• Not available when the option Digital interface [5] is not added
or when WIKA protocol is selected in Communication protocol
selection
• None :
Even :
Odd :
Even
Stop bit selection
• Selects the stop bit.
• Not available when the option Digital interface [5] is not added
or when WIKA protocol is selected in Communication protocol
selection
• Setting range: 1 or 2
1