11.2 Information on Foundation Fieldbus
Block diagram, measured value processing
The following illustration shows the Transducer Block and Function block in simplied form.
t
i
Simulate
ALARMS
OutPV
FIELD-VALUE
Mode & Status
PVF-TIME
LOW-CUTOFF
L-TYPE
OUT-SCALE
PV-SCALE
DIRECT
INDIRECT
INDIRECT
SQRT
bar % Lin%
Primary
Value
Secondary
Value 1
Secondary
Value 2
Sensor mounting
correction
AI 2
AI 3
AI 1
TB
Temperature
Sensor_Value
Linearizationadjustment
Channel = 1: Primary Value
Channel = 2: Secondary Value 1
Channel = 3: Secondary Value 2
Channel = 4: Temperature
Fig. 52: Transducer Block DPT10
TB Transducer Block
AI Function Block (AI =Analogue Input)
Diagram, adjustment
The following illustration shows the function of the adjustment:
Secondary_value_2
scale_in_0 scale_in_100 Sensor_value
0 %
100 %
bar
0,500-0,500
0
%
100
%
scale_in_100scale_in_0
Fig. 53: Adjustment DPT10
Parameter list
The following list contains the most important parameters and their meaning: